We report on a systematic, quantitative analysis of neuronal and inflammation markers across cortical depth. We implanted a total of 103 stiff silicon or flexible polyimide probes in 32 mice, varying their thicknesses and widths, which were either attached to the skull or not. A new, automated workflow to quantify immunohistochemical data examines: 1) the tissue loss caused by the implant, 2) the cortical neuronal density, and 3) the immune response expressed by astrocytic and microglial reaction.
Flexible polyimide probes exhibited a clear advantage, with fewer lesions and weaker immune responses than stiff silicon probes. Furthermore, we observed a weaker influence of the shank cross-section. A cortical depth profile of immune reactivity revealed focal reactions at the device entry points in the superficial cortex and at the cortex-white matter boundary. This study gives important insights on optimizing device design parameters as well as surgical insights for improved tissue integration of intracortical electrode arrays.
Recommended posters
POLYIMIDE-BASED MODULAR PROBE PLATFORM FOR MINIMALLY INVASIVE ACCESS TO DEEP NEURAL STRUCTURES
Levente Víg, Balázs Barkóczi, Marc Keller, Patrick Ruther, Domonkos Horváth, Klaudia Csikós, Rebeka Stelcz, Eszter Juhász, Dániel Hillier, István Ulbert, Richárd Fiáth
A SURFACE DRUG DELIVERY SYSTEM REDUCES THE FOREIGN BODY REACTION AGAINST INTRANEURAL ELECTRODES
Jose Manuel Crugeiras Pérez, Giulia Turrin, Chiara Bisquoli, Bruno Rodriguez-Meana, Claudio Trapella, Tomas Stieglitz, Yara Baslan, Hanna Karlsson-Fernberg, Fernand L. Narvaez-Chicaiza, Ruben López-Vales, Maria Asplund, Stefano Carli, Natalia Lago-Perez, Xavier Navarro
FLEXIBLE NANOPOROUS GRAPHENE MICROELECTRODE TECHNOLOGY FOR CHRONIC BIDIRECTIONAL NEURAL INTERFACING
Victoria Andino Pavlovsky, Eduard Masvidal Codina, Alexandros Katirtsidis, M. Victoria Puig, Jose A. Garrido
REUSABLE REDUCED-WEIGHT NEUROPIXELS IMPLANT FOR STABLE CHRONIC RECORDINGS IN SMALL ADULT MICE
Mengzhen Zhang, Alfredo Llorca, Patricia Maeso, Nathalie Rochefort
A ROBUST BRAIN IMPLANTATION METHOD FOR LONG-TERM WIRELESS ELECTROPHYSIOLOGICAL RECORDING IN FREELY BEHAVING MINIPIGS
Han-Chi Pan, Chia-Min Yeh, Yi-Chun Chou, Tzu-Min Huang, Jeng-Wei Yang, Da-Tong Ju, Hsin Chen
MULTIELECTRODE INTRACORTICAL MICROSTIMULATION ENABLED BY FLEXIBLE MULTISHANK PROBES DURING TWO-PHOTON CALCIUM IMAGING
Balazs Barkoczi, Eszter Nguyen, Csaba Horváth, Linda Judák, Balázs Rózsa, Frederik Ceyssens, István Ulbert, Lucia Wittner, Maarten Schelles, Richárd Fiáth